Instructions
RIGOL         Chapter 4 Functions of the Front Panel of EMI 
4-64              RSA3000E User Guide  
Trace+State 
Press Trace+State to enter the trace+state saving menu. The state of the 
instrument and the selected trace can be saved to the register or the file. 
1.  Save 
Press this key to save the current trace+state file in the default filename or 
user-defined filename. 
The default filename is determined by the following methods: 
  If you haven't input a filename by pressing the Save As key, its default 
filename is tracestate<n>.trs. Wherein, n refers to the file number of the 
filenames that are named based on the naming mode in the current file 
directory. The value of <n> is the existing maximum file number plus 1. 
  If you have pressed the Save As key to input the filename abc, then the 
file is saved to be "abc.trs". After that, press the Save key again, another 
file is saved automatically with the filename "abc1.trs". 
2.  Save As 
Press this key and then use the numeric keypad on the front panel to input the 
filename. Then, press Save to save the corresponding trace+state file. 
You can also touch the screen or use the mouse to tap or click the Save As 
menu, and then input the filename with the keypad. 
3.  Save From Trace 
Selects the trace to be saved. You can select any single trace from Trace1 to 
Trace3. 
4.  Register 1 through Register 16 
When any one of the items from Register 1 to Register 16 is selected, the 
current state of the instrument will be saved to the corresponding register. The 
register supports quick save and loading instrument state. The register menu 
displays the time for saving the instrument state. 
Measurement Data 
Press Meas Data to enter the measurement data saving menu. The selected 
measurement data type (e.g. trace, measurement results) can be saved to the 
specified file. The data will be saved in .csv format. They are separated with a 
comma, and this is convenient for you to use the Excel to analyze the data. 
1.  Save 
Press this key to save the currently selected type of measurement data in the 