Instructions
Chapter 1 Quick Start  RIGOL 
RSA3000E User Guide  1-13 
CAUTION 
To avoid damaging your hearing, please turn the volume down 
to zero first and then gradually turn the volume up after 
putting on the earphone. 
3.  Gen Output 50Ω 
The output of the tracking generator can be connected to a receiver through a 
cable with an N male connector. This function is only available for 
RSA3030E-TG/RSA3015E-TG. 
CAUTION 
To avoid damage to the tracking generator, the reverse power 
cannot exceed +10 dBm when the frequency is lower than 10 
MHz; the reverse power cannot exceed +20 dBm when the 
frequency is greater than 10 MHz. The reverse DC voltage 
should not exceed 50 V. 
4.  RF Input 50Ω 
The input terminal of the signal under test. [RF Input 50Ω] can be connected 
to the device under test (DUT) via a cable with an N male connector. 
CAUTION 
For the signal input to the RF input terminal, ensure that the 
DC voltage component and the maximum continuous power of 
the AC (RF) signal component do not exceed 50 V and +30 
dBm to avoid damaging the instrument. 