Chapter 4 Functions of the Front Panel of EMI  RIGOL 
RSA3000E User Guide  4-25 
  Pk Crit & LIM: performs the search and finds the peak that meets the peak 
criteria and limit line criteria. 
  SubRng & LIM: performs the search and finds the peak that meets the peak 
criteria for each subrange and limit line criteria. After this mode is selected, 
the whole span is divided into n subranges. Wherein, n is set by the 
Subrange Num menu. 
15. Peak Num State 
Enables or disables the setting of the number of peaks. 
16. Peak Num 
When the search criteria is "Peak Criteria" or "Pk Crit & LIM", set the maximum 
number of peaks searched. 
17. Subrange Num 
Sets the number of subranges for the signal search. 
Scan Table 
Sets the parameters for 10 ranges in the scan table. 
Note: The scan table currently displayed on the screen can be saved to the internal 
or external memory. If needed, you can recall it at any time. Press Save to save it 
according to descriptions in "Save". If a scan sequence is currently executed, then 
the Scan Table menu is grayed out and disabled. 
1.  Select Range 
Specifies the selected range. 10 ranges are available for you to choose. 
2.  Range 
Enables or disables the selected scan range. When enabled, the selected range 
will be chosen as a part of the measurement. Besides, you can tap the checkbox 
at the left side of the selected range in the pre-scan setting table. If the 
checkbox is highlighted and with a tick "√" being displayed, it indicates that the 
selected range is enabled. 
3.  Start Freq 
Sets the start frequency of the selected scan range. 
4.  Stop Freq 
Sets the stop frequency of the selected scan range. 
5.  Points 
Sets the number of scan points for the selected scan range. 