Instructions
Chapter 4 Functions of the Front Panel of EMI RIGOL
RSA3000E User Guide 4-23
CAvg Amp
indicates CISPR average detector amplitude
Displays the corresponding amplitude when the current trace
detector is CISPR average after the search operationis
executed. Displays the corresponding detection amplitude of
the final measurement detector 3 type after the final
measurement is performed. If there is no measurement data,
"--" is displayed.
Peak Lim1
△
Difference between positive peak detector amplitude and
limit1.
Only when the corresponding limit line is enabled and the final
measurement is performed, the difference between the
detector amplitude of the currently selected measurement
detector1 type and the limit line (1 to 6) is displayed. If there is
no measurement data, "--" is displayed.
QP Lim1
△
Difference between quasi-peak detector amplitude and limit1.
Only when the corresponding limit line is enabled and the final
measurement is performed, the difference between the
detector amplitude of the currently selected measurement
detector2 type and the limit line (1 to 6) is displayed. If there is
no measurement data, "--" is displayed.
CAvg Lim1
△
Difference between CISPR average detector amplitude and
limit1.
Only when the corresponding limit line is enabled and the final
measurement is performed, the difference between the
detector amplitude of the currently selected measurement
detector3 type and the limit line (1 to 6) is displayed. If there is
no measurement data, "--" is displayed.
Note: The signal table currently displayed on the screen can be saved to the internal
or external memory. If needed, you can recall it at any time. Press Save to save it
according to descriptions in "Save".
1. Select Signal
Selects signal n in the signal table.
2. Page Up
If the current signal table has several pages of data, press this key to go to the
previous page to view the data.
3. Page Down
If the current signal table has several pages of data, press this key to go to the
next page to view the data.
4. Signal->Meters
Modifies the frequency of the meter to the frequency of the selected signal.