Instructions
Chapter 4 Functions of the Front Panel of EMI  RIGOL 
RSA3000E User Guide  4-9 
Sweep and Function Settings 
BW 
RBW(Meter) 
Sets the resolution bandwidth (RBW) of the meter. You can use the numeric keys, the 
knob, or arrow keys to modify this parameter; you can also use the touch screen to 
modify the parameter.  
Table 4-10 RBW 
Parameter 
Remarks 
Default 
120 kHz 
Range 
100 Hz to 10 MHz
[1]
Unit 
GHz, MHz, kHz, Hz 
Knob Step 
at 1-3-10 step 
Left/Right Arrow Key Step 
Up/Down Arrow Key Step 
at 1-3-10 step 
Note
[1]
: When the EMC measurement standard is "None" and "Gauss" filter is selected, the RBW 
range is as above. When "EMI" filter is selected, resolution bandwidth can only be 200 Hz, 9 kHz, 
120 kHz, or 1 MHz. 
RBW Mode(Meter) 
Sets the RBW mode of the meter to "Manual" or "Auto". 
Remarks: 
  When you select "Manual", use the RBW(Meter) menu to set the RBW value. 
  When you select "Auto", use the RBW to automatically couple to the frequency 
of the meter. 
  When the EMC measurement standard is "None", the RBW Mode(Meter) 
menu is grayed out and disabled. You can manually set the RBW. When the EMC 
measurement standard is "CISPR", select the Auto RBW mode. 
Frequency (Meter) Fmet 
RBW 
Fmet < 150 kHz  
200 Hz 
150 kHz ≤ Fmet < 30 MHz 
9 kHz 
30 MHz ≤ Fmet < 1 GHz 
120 kHz 
Fmet ≥ 1 GHz 
1 MHz 