Instructions
RIGOL         Chapter 4 Functions of the Front Panel of EMI 
4-8            RSA3000E User Guide     
Scale/Div 
Sets the unit per division in the vertical axis.  
Remarks: 
  The amplitude range to be displayed can be adjusted by setting the scale. 
  The range of the signal amplitude that can be displayed is as follows: 
Minimum: reference level – 10 × current scale value 
Maximum: reference level 
  You can use the numeric keys, the knob, or arrow keys to modify this parameter; 
you can also use the touch screen to modify the parameter. 
Table 4-8 Scale/Div 
Parameter 
Remarks 
Default 
10 dB 
Range 
0.1 dB to 20 dB 
Unit 
dB 
Knob Step 
scale ≥ 1, step = 1 dB 
scale < 1, step = 0.1 dB 
Left/Right Arrow Key Step 
Up/Down Arrow Key Step 
at 1-2-5 step 
Ref Offset 
Adds an offset value to the reference level to compensate for gains or losses 
generated between the DUT and the analyzer input. 
Remarks: 
  The offset value does not affect the trace position, but will modify the reference 
level readout and the marker amplitude readout.  
  You can use the numeric keys, the knob, or arrow keys to modify this parameter; 
you can also use the touch screen to modify the parameter. 
Table 4-9 Reference Level Offset 
Parameter 
Remarks 
Default 
0 dB 
Range 
-300 dB to 300 dB 
Unit 
dB 
Knob Step 
level offset ≥ 1 or level offset ≤ -1: step = 1 dB; 
-1 < level offset < 1, step = 0.1 dB 
Left/Right Arrow Key Step 
Up/Down Arrow Key Step 
5 dB 