Instructions
RIGOL Chapter 2 Functions of the Front Panel of GPSA
2-92 RSA3000E User Guide
Preset
Recalls the preset setting and restores the system settings of the analyzer to a
specified status.
Remarks:
To set the preset type, press System Preset Preset Type to select
"Default" or any one of the items from "User1" to "User6".
Press Preset to recall the specified factory settings (as shown in the following
table, except the items specified in Note[3]) or user settings.
Parameter Name
GPSA Parameter Value
FREQ
Center Freq
1.5 GHz
Start Freq
0 Hz
Stop Freq
3 GHz
CF Step
Auto, 300 MHz
Freq Offset
0 Hz
Signal Track
Off
SPAN
Span
3 GHz
AMPT
Ref Level
0 dBm
Attenuation
Auto, 10 dB
RF Preamp
Off
Y Axis Unit
dBm
Scale Type
Log
Scale/Div
10 dB
Max Mixer Lvl
-10 dBm
Ref Offset
0 dB
BW
RBW
Auto, 3 MHz
SPAN/RBW Ratio
Auto, 106
VBW
Auto, 3 MHz
VBW/RBW Ratio
Auto, 1
Filter Type
Gaussian
Sweep
Sweep Points
801
Sweep Time
Auto, 1 ms
Sweep Time Mode
Continuous
Sweep Time Rule
Normal
Trigger
Trigger Source
Free Run
Trigger Hold-off
Off, 100 ms