Instructions
Chapter 2 Functions of the Front Panel of GPSA  RIGOL 
RSA3000E User Guide  2-77 
For details, refer to descriptions in "Parameter Setting". 
Table 2-57 N dB BW Parameter Settings 
Parameter 
Remarks 
Default 
-3.01 dB 
Range 
-140 dB to -0.01 dB 
Unit 
dB 
Knob Step 
1 dB 
Left/Right Arrow Key Step 
Up/Down Arrow Key Step 
1 dB 
Band Function 
The band function defines the corresponding parameters for a frequency band to 
measure the signal at the marker point. When the band function is enabled and 
"Detector Auto" is enabled, then the Detector Type is automatically modified to 
"Average (RMS)". 
1.  Noise 
When the noise measurement function is enabled, the measurement result of 
the Y-axis is the average noise level normalized to 1 Hz within the frequency 
band. 
2.  Band Power 
In non-zero span mode, it calculates the total power within a span. In zero span 
mode, it calculates the average power within a certain time range. 
3.  Band Density 
In non-zero span mode, the band density is the total power within the 
bandwidth to be measured divided by the measurement bandwidth. In zero 
span mode, the band density is the measured band power divided by Bn (Bn 
refers to the noise bandwidth of the RBW filter.) 
4.  Off 
Disables the band function. Disabling the band function will neither affect the 
frequency band parameters nor disable the marker. 