Instructions
RIGOL  Chapter 2 Functions of the Front Panel of GPSA 
2-64  RSA3000E User Guide 
Table 2-53 Average Count for TOI Measurement 
Parameter 
Remarks 
Default 
10 
Range 
1 to 1,000 
Unit 
None 
Knob Step 
1 
Left/Right Arrow Key Step 
Up/Down Arrow Key Step 
1 
2.  Avg Mode 
Sets the average mode to "Exponential" or "Repeat". The default average mode 
is "Exponential". 
  When "Exponential" is selected, the result is the exponential average of the 
measurement results obtained in the past N times (N is specified in "Avg 
Number"). 
  When "Repeat" is selected, the result is the arithmetic average of the 
measurement results obtained in the past N times (N is specified in "Avg 
Number"). 
3.  Avg State 
Selects whether to averaging the measurement results. By default, it is set to 
"On". 
4.  TOI Span 
This span is consistent with the span of the spectrum analyzer. It refers to the 
frequency range of the sweep. After it is set, the span of the spectrum analyzer 
will be modified accordingly. You can use the numeric keys, the knob, or arrow 
keys to modify this parameter; you can also use the touch screen to modify the 
parameter. 
Table 2-54 Span for TOI Measurement 
Parameter 
Remarks 
Default 
2 MHz 
Range 
10 Hz to 3 GHz 
Unit 
GHz, MHz, kHz, Hz 
Knob Step 
TOI distortion span/100, 
Min = 1 Hz 
Left/Right Arrow Key Step 
Up/Down Arrow Key Step 
at 1-1.5-2-3-5-7.5 step 
For details about the Auto Couple and Meas Preset menus, refer to their 
descriptions on Page 
2-45. 