Data Sheet

6
RIGOL
Level Display
Logarithmic Scale 1 dB to 200 dB
Linear Scale 0 to reference level
Number of Display Points 801
Number of Traces 6
Trace Detector
normal, pos-peak, neg-peak, sample, RMS average, voltage average, and
quasi-peak (Option RSA3000E-EMC)
Trace Function clear write, max hold, min hold, average, view, blank
Scale Unit dBm, dBmV, dBμV, nV, μV, mV, V, nW, μW, mW, W
Frequency Response
attenuation = 10 dB, relative to 50 MHz, 20
to 30
Preamp off 100 kHz to 3.0 GHz <0.7 dB, <0.5 dB (typical)
attenuation = 0 dB, relative to 50 MHz, 20
to 30
Preamp on 100 kHz to 3.0 GHz <1.0 dB, <0.5 dB (typical)