Data Sheet

11
RIGOL
EMI Mode (Option RSA3000E-EMI)
VSA Mode (Option RSA3000E-ASK/FSK)
EMI Resolution Bandwidth
Resolution Bandwidth (-3 dB) 100 Hz to 3 MHz, in 1-3-10 sequence
Resolution Bandwidth (-6 dB)
200 Hz, 9 kHz, 120 kHz
and 1 MHz
EMI Detector
Detector pos-peak, neg-peak, average, quasi-peak, CISPR average, RMS average
EMI Key Feature
Key Feature
CISPR 16-1-1 detectors
CISPR 16-1-1 bandwidths
log and linear display
signal table
scan table
simultaneous detectors
automatic limit testing
measure at marker
delta to limit
step and swept scans
report generation
Capture Oversampling
Capture Oversampling 4, 8, 16
Capture Length
Capture Oversampling = 4 Maximum 4096
Capture Oversampling = 8 Maximum 2048
Capture Oversampling = 16 Maximum 1024
Sample Rate
Maximum Sample Rate 12.8 MHz
Symbol Rate
Symbol Rate
depends on capture oversampling
= sample rate/capture oversampling, ≥1 kHz
Usable I/Q Bandwidth
Usable I/Q Bandwidth symbol rate × capture oversampling/1.28
Trigger Mode
Trigger Mode free run, external1, external2, power (time), and FMT
Modulation Format
FSK 2FSK, 4FSK, and 8FSK
ASK 2ASK and 4ASK
Filter Type
Measurement Filter Type No Filter, RRC, Gaussian, Rectangular, and User Dened
Reference Filter Type Raised Cosine, RRC, Gaussian, Rectangular, and User Dened
Predened standard
Cellular GSM, NADC, WCDMA, PDC, PHP (PHS)
Wireless Networking Bluetooth, WLAN (802.11b), ZigBee
Others TETRA, DECT, APCO-25
Measurement Uncertainty
Specications apply under the following conditions:
temperature from +20 °C to +30 °C
signal level ≥ –25 dBm
properly adjusted reference level
offset between device’s center frequency and signal’s center frequency smaller than 5 % of
symbol rate
Random data sequence
Capture oversampling is set to 4.
Residual Error for FSK
Test Signal
The reference lter is RRC with rolloff factor 0.22. The measurement lter is RRC with rolloff
factor 0.22. The FSK reference deviation is a quarter of the symbol rate. The result length is 150
symbols. The center frequency is 1 GHz.
Residual Frequency Error RMS
Symbol Rate
100 kHz < 2.8% (nominal)
500 kHz < 2.8% (nominal)