Instructions
Chapter 2 Command Systems for GPSA and RTSA Modes  RIGOL 
RSA3000E Programming Guide  2-63 
:DISPlay:WINDow:TRACe:Y:DLINe:STATe 
Syntax 
:DISPlay:WINDow:TRACe:Y:DLINe:STATe OFF|ON|0|1 
:DISPlay:WINDow:TRACe:Y:DLINe:STATe? 
Description
Turns on or off the display line. 
Queries the status of the display line. 
Parameter
Name 
Type 
Range 
Default 
—— 
Bool 
OFF|ON|0|1 
OFF|0 
Return Format
The query returns 0 or 1. 
Example
The following command enables the display line.   
:DISPlay:WINdow:TRACe:Y:DLINe:STATe ON or :DISPlay:WINdow:TRACe:Y:DLINe:STATe 1 
The following query returns 1.  
:DISPlay:WINdow:TRACe:Y:DLINe:STATe? 
:DISPlay:WINDow:TRACe:Y[:SCALe]:NRLevel 
Syntax 
:DISPlay:WINDow:TRACe:Y[:SCALe]:NRLevel <rel_ampl> 
:DISPlay:WINDow:TRACe:Y[:SCALe]:NRLevel?
Description
Sets the reference level of normalization. 
Queries the reference level of normalization. 
Parameter
Name 
Type 
Range 
Default 
<rel_ampl> 
Consecutive Real Number 
-200 dB to 200 dB 
10 dB 
Remarks
This command is only available for RSA3030E-TG/RSA3015E-TG working in GPSA mode. 
This command is only valid when the tracking generator function is enabled.
Return Format
The query returns the reference level of normalization in scientific notation. 
Example
The following command sets the reference level of normalization to -20 dB.  
:DISPlay:WINDow:TRACe:Y:SCALe:NRLevel -20 
The following query returns -2.000000e+01.   
:DISPlay:WINDow:TRACe:Y:SCALe:NRLevel? 