Instructions
Chapter 2 Command Systems for GPSA and RTSA Modes  RIGOL 
RSA3000E Programming Guide  2-51 
:DISPlay:PVTime:WINDow:TRACe:X[:SCALe]:COUPle 
Syntax 
:DISPlay:PVTime:WINDow:TRACe:X[:SCALe]:COUPle OFF|ON|0|1 
:DISPlay:PVTime:WINDow:TRACe:X[:SCALe]:COUPle? 
Description 
Enables or disables the auto scale function for the horizontal axis in the PvT view. 
Queries the status of the auto scale function for the horizontal axis in the PvT view. 
Parameter 
Name 
Type 
Range 
Default 
-- 
Bool 
OFF|ON|0|1 
ON|1 
Remarks 
This command is only valid when the PvT view in RTSA mode is selected. 
Return Format
The query returns 1 or 0. 
Example
The following command enables the auto scale function for the horizontal axis in the PvT view. 
:DISPlay:PVTime:WINDow:TRACe:X:SCALe:COUPle ON 
or :DISPlay:PVTime:WINDow:TRACe:X:SCALe:COUPle 1 
The following query returns 1. 
:DISPlay:PVTime:WINDow:TRACe:X:SCALe:COUPle?
:DISPlay:PVTime:WINDow:TRACe:X[:SCALe]:PDIVision 
Syntax 
:DISPlay:PVTime:WINDow:TRACe:X[:SCALe]:PDIVision <time> 
:DISPlay:PVTime:WINDow:TRACe:X[:SCALe]:PDIVision? 
Description 
Sets the unit per division in the horizontal axis of the PvT view. 
Queries the unit per division in the horizontal axis of the PvT view. 
Parameter 
Name 
Type 
Range 
Default 
<time>  Consecutive Real Number  20 μs to 4 s 
acquisition 
time/10 
Remarks 
This command is only valid when the PvT view in RTSA mode is selected. 
Return Format
The query returns the X-axis scale value in scientific notation. 
Example
The following command sets the horizontal scale in the PvT view to 40 μs. 
:DISPlay:PVTime:WINDow:TRACe:X:SCALe:PDIVision 0.00004 