Instructions
Chapter 2 Command Systems for GPSA and RTSA Modes RIGOL
RSA3000E Programming Guide 2-31
:CALCulate:MARKer<n>[:SET]:RLEVel
Syntax
:CALCulate:MARKer<n>[:SET]:RLEVel
Description
Sets the amplitude of the specified marker to the reference level of the analyzer.
Parameter
Name
Type
Range
Default
<n>
Discrete
1|2|3|4|5|6|7|8
——
Remarks
If the marker mode of the specified marker is Position or Fixed, the reference level will be set to the
amplitude of the marker.
If the specified marker mode is Delta and the current marker is the reference marker, then the reference
level is set to the amplitude of the reference marker; if the current marker is the Delta marker, then the
reference level is set to the amplitude of the Delta marker.
Example
The following command sets the amplitude of Marker 2 (Position) to the reference level of the analyzer.
:CALCulate:MARKer2:SET:RLEVel
:CALCulate:MARKer<n>[:SET]:STARt
Syntax
:CALCulate:MARKer<n>[:SET]:STARt
Description
Sets the frequency of the specified marker to the start frequency of the analyzer.
Parameter
Name
Type
Range
Default
<n>
Discrete
1|2|3|4|5|6|7|8
——
Remarks
If the marker mode of the specified marker is Position or Fixed, the start frequency will be set to the
frequency of the marker.
If the specified marker mode is Delta, the start frequency will be set to the frequency of the Delta marker.
This function is invalid in zero span.
This command is invalid for the PvT measurement mode in RTSA.
Example
The following command sets the frequency of Marker 3 (Position) to the start frequency of the analyzer.
:CALCulate:MARKer3:SET:STARt