Instructions
Chapter 6 Appendix  RIGOL 
RSA3000E Programming Guide  6-1 
Chapter 6  Appendix 
Appendix A: Default Settings 
Sending the 
*RST command can restore the instrument to default settings, as shown in the table below. 
Parameter Name 
GPSA Parameter 
Value 
RTSA Parameter Value   
Non-PvT 
PvT 
FREQ 
Center Freq 
1.5 GHz 
1.5 GHz 
Start Freq 
0 Hz 
1.495 GHz 
Stop Freq 
3 GHz 
1.505 GHz 
CF Step 
Auto, 300 MHz 
Auto, 1 MHz 
Freq Offset 
0 Hz 
0 Hz 
Signal Track 
Off 
-- 
SPAN 
Span 
3 GHz 
10 MHz 
Ref Value 
-- 
-- 
0 us 
Scale/Div 
-- 
-- 
3.1946 ms 
Ref Position 
-- 
-- 
Left 
Auto Scale 
-- 
-- 
Auto 
AMPT 
Ref Level 
0 dBm 
0 dBm 
Attenuation 
Auto, 10 dB 
Auto, 10 dB 
RF Preamp 
Off 
Off 
Y Axis Unit 
dBm 
dBm 
Scale Type 
Log 
Log 
Scale/Div 
10 dB 
10 dB 
Max Mixer Lvl 
-10 dBm 
-10 dBm 
Ref Offset 
0 dB 
0 dB 
BW 
RBW 
Auto, 3 MHz 
Auto, 50.228 kHz 
(RBW2) 
-- 
SPAN/RBW Ratio 
Auto, 106 
-- 
-- 
VBW 
Auto, 3 MHz 
-- 
-- 
VBW/RBW Ratio 
Auto, 1 
-- 
-- 
Filter Type 
Gaussian 
Kaiser 
-- 
Sweep 
Sweep Points 
801 
-- 
Sweep Time 
Auto, 1 ms 
-- 
Acq Time 
-- 
Auto, 32 ms 
Auto, 31.9460 ms 
Sweep Time Mode 
Continuous 
Continuous 
Sweep Time Rule 
Normal 
-- 
Trigger 
Trigger Source 
Free Run 
Free Run 
Trigger Holdoff 
Off, 100 ms 
Off, 100 ms 
Auto Trig 
Off, 100 ms 
Off, 100 ms 
Slope 
Positive 
Positive 
Trig Delay 
Off, 1 us 
Off, 1 us 
Trigger Level 
-25 dBm 
-- 
Acq/Trigger 
-- 
1 
Power 
-- 
0 dB 
Mask Type 
-- 
Upper 
Trigger Mask 
-- 
Upper 