Instructions
RIGOL  Chapter 4 Command System for VSA Mode 
4-56  RSA3000E Programming Guide 
[:SENSe]:FREQuency:STOP 
Syntax 
[:SENSe]:FREQuency:STOP <freq>  
[:SENSe]:FREQuency:STOP? 
Description
Sets the stop frequency.  
Queries the stop frequency.
Parameter 
Name 
Type 
Range 
Default 
<freq>  
Consecutive Real Number 
3.125 kHz to Fmax 
center frequency + span/2 
Return Format
The query returns the stop frequency in scientific notation. The unit is Hz. 
Example
The following command sets the stop frequency to 10 MHz.  
:SENSe:FREQuency:STOP 10000000 
The following query returns 1.000000000e+07.   
:SENSe:FREQuency:STOP? 
[:SENSe]:POWer[:RF]:RANGe 
Syntax 
[:SENSe]:POWer[:RF]:RANGe <real>   
[:SENSe]:POWer[:RF]:RANGe? 
Description
Sets the amplitude of the largest sinusoidal input signal without being clipped by the IF ADC. 
Queries the amplitude of the largest sinusoidal input signal without being clipped by the IF ADC. 
Parameter 
Name 
Type 
Range 
Default 
<real>  
Integer 
-15 dBm to 25 dBm 
20 dBm 
Return Format
The query returns the signal amplitude in scientific notation. The unit is dBm. 
Example
The following command sets the amplitude of the largest sinusoidal input signal without being clipped by 
the IF ADC to 20 dBm.   
:SENSe:POWer:RF:RANGe 20   
The following query returns 2.000000000e+01.  
:SENSe:POWer:RF:RANGe? 