Instructions
Chapter 4 Command System for VSA Mode  RIGOL 
RSA3000E Programming Guide  4-49 
Parameter 
Name 
Type 
Range 
Default 
<frequency> 
Consecutive Real Number 
Refer to "Remarks" 
1 MHz 
Remarks 
The minimum symbol rate is 1 kHz. 
The maximum symbol rate SRmax = SPmax × 1.28/(points/symbol). Wherein, SPmax is the maximum 
span.  
Return Format
The query returns the symbol rate in scientific notation. 
Example
The following command sets the symbol rate to 1 MHz.  
:SENSe:DDEMod:SRATe 1000000 
The following query returns 1.000000000e+06.   
:SENSe:DDEMod:SRATe? 
[:SENSe]:DDEMod:STANdard:PRESet 
Syntax 
[:SENSe]:DDEMod:STANdard:PRESet BLUETOOTH|DECT|APCO 
Description
Sets the preset standard of the digital demodulation. 
Parameter 
Name 
Type 
Range 
Default 
-- 
Keyword 
BLUETOOTH|DECT|APCO 
-- 
Return Format 
The query returns BLUETOOTH, DECT, or APCO. 
Example
The following command sets the preset standard of the digital demodulation to DECT.  
:SENSe:DDEMod:STANdard:PRESet DECT 
[:SENSe]:DDEMod:SWEep:POINts 
Syntax 
[:SENSe]:DDEMod:SWEep:POINts <integer> 
[:SENSe]:DDEMod:SWEep:POINts? 
Description
Sets the number of symbols in digital demodulation. 
Queries the number of symbols in digital demodulation. 
Parameter 
Name 
Type 
Range 
Default 
<integer> 
Integer 
Refer to "Remarks" 
50 