Instructions
Chapter 4 Command System for VSA Mode  RIGOL 
RSA3000E Programming Guide  4-33 
:INSTrument Commands 
Command List: 
  :INSTrument:COUPle:FREQuency:CENTer 
  :INSTrument:DEFault 
  :INSTrument:NSELect 
  :INSTrument[:SELect] 
:INSTrument:COUPle:FREQuency:CENTer 
Syntax 
:INSTrument:COUPle:FREQuency:CENTer ALL|NONE 
:INSTrument:COUPle:FREQuency:CENTer? 
Description 
Turns on or off the global center frequency of the instrument. 
Queries the setting status of the global center frequency of the instrument. 
Parameter 
Name 
Type 
Range 
Default 
-- 
Keyword 
ALL|NONE 
NONE 
Remarks 
NONE: turns off the global center frequency. 
ALL: turns on the global center frequency. 
If you execute this command in any mode, the center frequency of the current mode is set to the global 
center frequency. Adjusting the center frequency in a mode, while the global center frequency is on, will 
modify the global center frequency. 
Return Format 
The query returns ALL or NONE. 
Example 
The following command enables the global center frequency of the instrument. 
:INSTrument:COUPle:FREQuency:CENTer ALL 
The following query returns ALL. 
:INSTrument:COUPle:FREQuency:CENTer? 
:INSTrument:DEFault 
Syntax 
:INSTrument:DEFault 
Description 
Resets the parameters of the current mode to the factory default settings. 