Instructions
RIGOL  Chapter 4 Command System for VSA Mode 
4-32  RSA3000E Programming Guide 
:INITiate Command 
Command List: 
  :INITiate:BER:RESTart 
  :INITiate:CONTinuous 
:INITiate:BER:RESTart 
Syntax 
:INITiate:BER:RESTart 
Description 
Restarts the BER test. 
Remarks 
This command is valide only when the BER Test measurement function is selected. 
:INITiate:CONTinuous 
Syntax 
:INITiate:CONTinuous OFF|ON|0|1 
:INITiate:CONTinuous?
Description 
Selects continuous (ON|1) or single (OFF|0) sweep mode. 
Queries the current sweep mode. 
Parameter 
Name 
Type 
Range 
Default 
-- 
Bool 
OFF|ON|0|1 
ON|1 
Return Format
The query returns 0 or 1. 
Example
The following command sets the instrument to sweep continuously.  
:INITiate:CONTinuous ON or :INITiate:CONTinuous 1 
The following query returns 1.  
:INITiate:CONTinuous? 










