Instructions
RIGOL  Chapter 4 Command System for VSA Mode 
4-14  RSA3000E Programming Guide 
Parameter 
Name 
Type 
Range 
Default 
<n>  
Discrete 
1|2|3|4|5|6|7|8 
-- 
Remarks
If the marker mode of the specified marker is Position or Fixed, the reference level will be set to the 
amplitude of the marker.
If the specified marker mode is Delta and the current marker is the reference marker, then the reference 
level is set to the amplitude of the reference marker; if the current marker is the Delta marker, then the 
reference level is set to the amplitude of the Delta marker. 
Example 
The following command sets the reference level of the analyzer to the amplitude of Marker 2 (Position). 
:CALCulate:MARKer2:SET:RLEVel 
:CALCulate:MARKer<n>[:SET]:STARt 
Syntax 
:CALCulate:MARKer<n>[:SET]:STARt
Description
Sets the start frequency of the analyzer to the frequency of the specified marker. 
Parameter 
Name 
Type 
Range 
Default 
<n>  
Discrete 
1|2|3|4|5|6|7|8 
-- 
Remarks
If the marker mode of the specified marker is Position or Fixed, the start frequency will be set to the 
frequency of the marker. 
If the specified marker mode is Delta, the start frequency will be set to the frequency of the Delta marker. 
Example
The following command sets the start frequency of the analyzer to the frequency of Marker 3 (Position). 
:CALCulate:MARKer3:SET:STARt 
:CALCulate:MARKer<n>[:SET]:STEP 
Syntax 
:CALCulate:MARKer<n>[:SET]:STEP 
Description
Sets the center frequency step of the analyzer to the frequency of the specified marker. 
Parameter 
Name 
Type 
Range 
Default 
<n>  
Discrete 
1|2|3|4|5|6|7|8 
-- 
Remarks
If the marker mode of the specified marker is Position or Fixed, the center frequency step will be set to the 
frequency of the marker. 
If the specified marker mode is Delta, the center frequency step will be set to the frequency difference 










