Instructions

RIGOL Chapter 4 Command System for VSA Mode
4-14 RSA3000E Programming Guide
Parameter
Name
Type
Range
Default
<n>
Discrete
1|2|3|4|5|6|7|8
--
Remarks
If the marker mode of the specified marker is Position or Fixed, the reference level will be set to the
amplitude of the marker.
If the specified marker mode is Delta and the current marker is the reference marker, then the reference
level is set to the amplitude of the reference marker; if the current marker is the Delta marker, then the
reference level is set to the amplitude of the Delta marker.
Example
The following command sets the reference level of the analyzer to the amplitude of Marker 2 (Position).
:CALCulate:MARKer2:SET:RLEVel
:CALCulate:MARKer<n>[:SET]:STARt
Syntax
:CALCulate:MARKer<n>[:SET]:STARt
Description
Sets the start frequency of the analyzer to the frequency of the specified marker.
Parameter
Name
Type
Range
Default
<n>
Discrete
1|2|3|4|5|6|7|8
--
Remarks
If the marker mode of the specified marker is Position or Fixed, the start frequency will be set to the
frequency of the marker.
If the specified marker mode is Delta, the start frequency will be set to the frequency of the Delta marker.
Example
The following command sets the start frequency of the analyzer to the frequency of Marker 3 (Position).
:CALCulate:MARKer3:SET:STARt
:CALCulate:MARKer<n>[:SET]:STEP
Syntax
:CALCulate:MARKer<n>[:SET]:STEP
Description
Sets the center frequency step of the analyzer to the frequency of the specified marker.
Parameter
Name
Type
Range
Default
<n>
Discrete
1|2|3|4|5|6|7|8
--
Remarks
If the marker mode of the specified marker is Position or Fixed, the center frequency step will be set to the
frequency of the marker.
If the specified marker mode is Delta, the center frequency step will be set to the frequency difference