Instructions

RIGOL Chapter 3 Command System for EMI Mode
3-86 RSA3000E Programming Guide
[:SENSe]:FSCan:TRACe:TYPE
Syntax
[:SENSe]:FSCan:TRACe:TYPE WRITe|AVERage|MAXHold|MINHold
[:SENSe]:FSCan:TRACe:TYPE?
Description
Sets the type of the trace.
Queries the type of the trace.
Parameter
Name
Type
Range
Default
--
Keyword
WRITe|AVERage|MAXHold|MINHold
WRITe
Remarks
WRITe: indicates clear/write.
AVERage: indicates average.
MAXHold: indicates the maximum hold.
MINHold: indicates the minimum hold.
Return Format
The query returns WRIT, AVER, MAXH, or MINH.
Example
The following command sets the type of the current trace to MAXHold.
:SENSe:FSCan:TRACe:TYPE MAXHold
The following query returns MAXH.
:SENSe:FSCan:TRACe:TYPE?
[:SENSe]:METer:DETector:DWELl
Syntax
[:SENSe]:METer:DETector:DWELl <time>
[:SENSe]:METer:DETector:DWELl?
Description
Sets the dwell time of the meter detector.
Queries the dwell time of meter detector.
Parameter
Name
Type
Range
Default
<time>
Consectutive Real Number
10 ms to 100 s
10 ms
Return Format
The query returns the dwell time in scientific notation. The unit is s.
Example
The following command sets the dwell time of the meter detector to 0.1 s.
:SENSe:METer:DETector:DWELl 0.1
The following query returns 1.000000000e-01.
:SENSe:METer:DETector:DWELl?