Instructions
Chapter 2 Command Systems for GPSA and RTSA Modes  RIGOL 
RSA3000E Programming Guide  2-7 
:CALCulate:LLINe<n>:AMPLitude:CMODe:RELative 
Syntax 
:CALCulate:LLINe<n>:AMPLitude:CMODe:RELative OFF|ON|0|1 
:CALCulate:LLINe<n>:AMPLitude:CMODe:RELative? 
Description 
Enables or disables the coupling functions between the selected limit line data points and the reference 
level. 
Queries the status of the coupling functions between the selected limit line data points and the reference 
level. 
Parameter 
Name 
Type 
Range 
Default 
<n> 
Discrete 
1|2|3|4|5|6 
——
—— 
Bool 
OFF|ON|0|1 
OFF|0 
Remarks 
OFF|0: disables the coupling function. That is, "Fixed" is selected under Y to Ref, and the amplitude of the 
current limit line editing point is not affected by the reference level.  
ON|1: enables the coupling function. That is, "Relative" is selected under Y to Ref, and the amplitude of 
the current limit line editing point is the difference between the current point with the current reference 
level. At this time, if the reference level changes, the position of the current editing point will move up and 
down with it. 
Return Format 
The query returns 0 or 1.
Example 
The following command disables the coupling functions between the selected limit line data points and the 
reference level.  
:CALCulate:LLINe2:AMPLitude:CMODe:RELative OFF or :CALCulate:LLINe2:AMPLitude:CMODe:RELative 0 
The following query returns 0. 
:CALCulate:LLINe2:AMPLitude:CMODe:RELative? 
:CALCulate:LLINe<n>:BUILd 
Syntax 
:CALCulate:LLINe<n>:BUILd TRACE1|TRACE2|TRACE3|TRACE4|TRACE5|TRACE6 
Description 
Builds the limit line from the selected trace. 
Parameter 
Name 
Type 
Range 
Default 
<n> 
Discrete 
1|2|3|4|5|6 
—— 
—— 
Keyword 
TRACE1|TRACE2|TRACE3|TRACE4|TRACE5|TRACE6 
—— 
Example 
The following command builds Limit Line 2 from Trace2. 
:CALCulate:LLINe2:BUILd TRACE2 