Instructions
RIGOL Chapter 3 Command System for EMI Mode
3-76 RSA3000E Programming Guide
[:SENSe]:FSCan:SCAN:ABORT
Syntax
[:SENSe]:FSCan:SCAN:ABORT
Description
Stops the current scan sequence.
[:SENSe]:FSCan:SCAN<n>:BANDwidth|BWIDth:EMIFilter:STATe
Syntax
[:SENSe]:FSCan:SCAN<n>:BANDwidth|BWIDth:EMIFilter:STATe OFF|ON|0|1
[:SENSe]:FSCan:SCAN<n>:BANDwidth|BWIDth:EMIFilter:STATe?
Description
Enables or disables the EMI filter within the specified range.
Queries the status of the EMI filter within the specified range.
Parameter
Name
Type
Range
Default
<n>
Discrete
1|2|3|4|5|6|7|8|9|10
3|4|5
--
Bool
OFF|ON|0|1
ON|1
Remarks
ON: selects the EMI filter (-6 dB bandwidth).
OFF: selects the Gaussian filter (-3 dB bandwidth).
This command is valid only when the EMC measurement standard is "None".
Return Format
The query returns 0 or 1.
Example
The following command selects the filter type within Range 2 to EMI filter.
:SENSe:FSCan:SCAN2:BANDwidth:EMIFilter:STATe ON
or :SENSe:FSCan:SCAN2:BANDwidth:EMIFilter:STATe 1
:SENSe:FSCan:SCAN2:BWIDth:EMIFilter:STATe ON or :SENSe:FSCan:SCAN2:BWIDth:EMIFilter:STATe 1
The following query returns 1.
:SENSe:FSCan:SCAN2:BANDwidth:EMIFilter:STATe?
:SENSe:FSCan:SCAN2:BWIDth:EMIFilter:STATe?
[:SENSe]:FSCan:SCAN<n>:BANDwidth[:RESolution]
Syntax
[:SENSe]:FSCan:SCAN<n>:BANDwidth[:RESolution] <freq>
[:SENSe]:FSCan:SCAN<n>:BANDwidth[:RESolution]?
Description
Sets the resolution bandwidth (RBW) for the specified range.
Queries the RBW for the specified range.