Instructions
Chapter 3 Command System for EMI Mode  RIGOL 
RSA3000E Programming Guide  3-75 
[:SENSe]:FSCan:FINal:DETector<n>:LDELta 
Syntax 
[:SENSe]:FSCan:FINal:DETector<n>:LDELta <limit_Num> 
[:SENSe]:FSCan:FINal:DETector<n>:LDELta? 
Description 
Sets the limit line used by the specified detector in the final measurement. 
Queries the limit line used by the specified detector in the final measurement. 
Parameter 
Name 
Type 
Range 
Default 
<n> 
Discrete 
1|2|3 
1 
<limit_Num> 
Discrete 
1|2|3|4|5|6 
1 
Return Format 
The query returns the limit line used by the specified detector. 
Example 
The following command sets the limit line used by Detector 1 in the final measurement to Limit Line 2. 
:SENSe:FSCan:FINal:DETector1:LDELta 2 
The following query returns 2. 
:SENSe:FSCan:FINal:DETector1:LDELta? 
[:SENSe]:FSCan:MARKer:COUPle:METer 
Syntax 
[:SENSe]:FSCan:MARKer:COUPle:METer OFF|ON|0|1 
[:SENSe]:FSCan:MARKer:COUPle:METer? 
Description 
Enables or disables the coupling function between the meter and the marker. 
Queries the status of the coupling function between the meter and the marker. 
Parameter 
Name 
Type 
Range 
Default 
-- 
Bool 
OFF|ON|0|1 
OFF|0 
Return Format 
The query returns 0 or 1. 
Example 
The following command enables the coupling function between the meter and the marker. 
:SENSe:FSCan:MARKer:COUPle:METer ON or :SENSe:FSCan:MARKer:COUPle:METer 1 
The following query returns 1. 
:SENSe:FSCan:MARKer:COUPle:METer? 