Instructions
Chapter 3 Command System for EMI Mode RIGOL
RSA3000E Programming Guide 3-69
Example
The following command enables the Detector Auto function for Trace 1.
:SENSe:DETector:TRACe1:AUTO ON or :SENSe:DETector:TRACe1:AUTO 1
The following query returns 1.
:SENSe:DETector:TRACe1:AUTO?
[:SENSe]:EMC:STANdard[:SELect]
Syntax
[:SENSe]:EMC:STANdard[:SELect] NONE|CISPr
[:SENSe]:EMC:STANdard[:SELect]?
Description
Sets the EMC standard.
Queries the EMC standard.
Parameter
Name
Type
Range
Default
--
Discrete
NONE|CISPr
CISPr
Return Format
The query returns NONE or CISP.
Example
The following command sets the EMC standard to CISPr.
:SENSe:EMC:STANdard:SELect CISPr
The following query returns CISP.
:SENSe:EMC:STANdard:SELect?
[:SENSe]:FREQuency:CENTer
Syntax
[:SENSe]:FREQuency:CENTer <freq>
[:SENSe]:FREQuency:CENTer?
Description
Sets the frequency of the meter.
Queries the frequency of the meter.
Parameter
Name
Type
Range
Default
<freq>
Consecutive Real Number
0 Hz to Fmax
[1]
515 MHz
Note
[1]
: The maximum measurement frequency Fmax is determined by the instrument model. For RSA3030E, Fmax is
3 GHz; for RSA3015E, Fmax is 1.5 GHz.
Return Format
The query returns the frequency of the meter in scientific notation. The unit is Hz.
Example
The following command sets the frequency of the meter to 1 MHz.