Instructions

RIGOL Chapter 3 Command System for EMI Mode
3-48 RSA3000E Programming Guide
Parameter
Name
Type
Range
Default
--
Bool
OFF|ON|0|1
ON|1
Return Format
The query returns 0 or 1.
Example
The following command sets the measurement mode of the meter to CONTinuous.
:INITiate1:CONTinuous ON or :INITiate1:CONTinuous 1
The following query returns 1.
:INITiate1:CONTinuous?
:INITiate2:CONTinuous
Syntax
:INITiate2:CONTinuous OFF|ON|0|1
:INITiate2:CONTinuous?
Description
Selects continuous (ON|1) or single (OFF|0) sweep mode.
Queries the current sweep mode.
Parameter
Name
Type
Range
Default
--
Bool
OFF|ON|0|1
ON|1
Return Format
The query returns 0 or 1.
Example
The following command sets the instrument to sweep continuously.
:INITiate2:CONTinuous ON or :INITiate:CONTinuous 1
The following query returns 1.
:INITiate2:CONTinuous?
:INITiate2:IMMediate
:INITiate2:RESTart
Syntax
:INITiate2:IMMediate
:INITiate2:RESTart
Description
Starts the scan, search, or the final measurement based on the selected scan sequence.