Instructions
RIGOL  Chapter 3 Command System for EMI Mode 
3-28  RSA3000E Programming Guide 
Parameter 
Name 
Type 
Range 
Default 
<n>  
Discrete 
1|2|3|4|5|6|7|8 
-- 
:CALCulate:METer:LIMit:ALL:FAIL? 
Syntax 
:CALCulate:METer:LIMit:ALL:FAIL? 
Description 
Queries the measurement results of all the limit lines and their associated meters. 
Remarks 
If no limit line is set for the meter, the query returns 0. 
It is only applicable to the meter whose limit lines are enabled. If one limit line is disabled, then it will not be 
tested. 
Return Format 
The query returns 0 (indicating that all the meters are within the limit, and the test is successful) or 1 
(indicating that the test is failed). 
:CALCulate:METer:POWer[:CURRent]? 
Syntax 
:CALCulate:METer:POWer[:CURRent]? 
Description 
Queries the current power of the meter. 
Remarks
If the meter filter is EMI filter and RBW is 1 MHz, and the meter detector type is quasi peak, the query of the 
corresponding meter power value returns -1.700000000e+02. 
If the meter filter is Gauss filter, and the meter detector type is quasi peak, CISPR average, or RMS average, 
the query of the corresponding meter power value returns -1.700000000e+02. 
The query returns the minimum data value (-1.700000000e+02) if the meter is disabled. 
Return Format
The query returns the current power values of three meters in scientific notation. The values are separated 
by commas.   
:CALCulate:METer:POWer:PEAK? 
Syntax 
:CALCulate:METer:POWer:PEAK? 
Description 
Queries the peak power of the meter. 
Remarks
The query returns the minimum data value if the meter is disabled. 