Instructions

RIGOL Chapter 3 Command System for EMI Mode
3-28 RSA3000E Programming Guide
Parameter
Name
Type
Range
Default
<n>
Discrete
1|2|3|4|5|6|7|8
--
:CALCulate:METer:LIMit:ALL:FAIL?
Syntax
:CALCulate:METer:LIMit:ALL:FAIL?
Description
Queries the measurement results of all the limit lines and their associated meters.
Remarks
If no limit line is set for the meter, the query returns 0.
It is only applicable to the meter whose limit lines are enabled. If one limit line is disabled, then it will not be
tested.
Return Format
The query returns 0 (indicating that all the meters are within the limit, and the test is successful) or 1
(indicating that the test is failed).
:CALCulate:METer:POWer[:CURRent]?
Syntax
:CALCulate:METer:POWer[:CURRent]?
Description
Queries the current power of the meter.
Remarks
If the meter filter is EMI filter and RBW is 1 MHz, and the meter detector type is quasi peak, the query of the
corresponding meter power value returns -1.700000000e+02.
If the meter filter is Gauss filter, and the meter detector type is quasi peak, CISPR average, or RMS average,
the query of the corresponding meter power value returns -1.700000000e+02.
The query returns the minimum data value (-1.700000000e+02) if the meter is disabled.
Return Format
The query returns the current power values of three meters in scientific notation. The values are separated
by commas.
:CALCulate:METer:POWer:PEAK?
Syntax
:CALCulate:METer:POWer:PEAK?
Description
Queries the peak power of the meter.
Remarks
The query returns the minimum data value if the meter is disabled.