Instructions

RIGOL Chapter 3 Command System for EMI Mode
3-18 RSA3000E Programming Guide
:CALCulate:FSCan:SLISt:ZOOM:OUT
Syntax
:CALCulate:FSCan:SLISt:ZOOM:OUT
Description
Returns the coordinate display of the current signal to its previous display set before the Signal Zoom.
:CALCulate:LLINe:ALL:DELete
Syntax
:CALCulate:LLINe:ALL:DELete
Description
Deletes all the limit line data.
:CALCulate:LLINe:TEST
Syntax
:CALCulate:LLINe:TEST OFF|ON|0|1
:CALCulate:LLINe:TEST?
Description
Enables or disables the limit line test function.
Queries the setting state of the limit line test function.
Parameter
Name
Type
Range
Default
--
Bool
OFF|ON|0|1
ON|1
Return Format
The query returns 0 or 1.
Example
The following command disables the limit line test function.
:CALCulate:LLINe:TEST OFF or :CALCulate:LLINe:TEST 0
The following query returns 0.
:CALCulate:LLINe:TEST?
:CALCulate:LLINe<n>:AMPLitude:CMODe:RELative
Syntax
:CALCulate:LLINe<n>:AMPLitude:CMODe:RELative OFF|ON|0|1
:CALCulate:LLINe<n>:AMPLitude:CMODe:RELative?
Description
Enables or disables the coupling functions between the selected limit line data points and the reference
level.
Queries the setting state of the coupling functions between the selected limit line data points and the