Instructions
RIGOL  Chapter 3 Command System for EMI Mode 
3-14  RSA3000E Programming Guide 
Parameter 
Name 
Type 
Range 
Default 
<integer> 
Integer 
1 to 2,000 
-- 
Example 
The following command clears the mark of Signal 20 in the signal table. 
:CALCulate:FSCan:SLISt:MARK:CLEar:SIGNal 20 
:CALCulate:FSCan:SLISt:MARK:SIGNal 
Syntax 
:CALCulate:FSCan:SLISt:MARK:SIGNal <integer> 
Description 
Marks the specified signal. 
Parameter
Name 
Type 
Range 
Default 
<integer> 
Integer 
1 to 2,000 
-- 
Example
The following command marks Signal 20 in the signal table.  
:CALCulate:FSCan:SLISt:MARK:SIGNal 20 
:CALCulate:FSCan:SLISt:REPLace:METer 
Syntax 
:CALCulate:FSCan:SLISt:REPLace:METer <integer>
Description
Replaces the frequency of the currently selected signal with the currently set frequency of the meter. 
Parameter 
Name 
Type 
Range 
Default 
<integer> 
Integer 
1 to 2,000 
-- 
Example 
The following command replaces the frequency of Signal 20 with the frequency of the meter. 
:CALCulate:FSCan:SLISt:REPLace:METer 20 
:CALCulate:FSCan:SLISt:SET:METer 
Syntax 
:CALCulate:FSCan:SLISt:SET:METer <integer> 
Description 
Sets the frequency of the meter to the frequency of the specified signal. 