Instructions

RIGOL Chapter 3 Command System for EMI Mode
3-14 RSA3000E Programming Guide
Parameter
Name
Type
Range
Default
<integer>
Integer
1 to 2,000
--
Example
The following command clears the mark of Signal 20 in the signal table.
:CALCulate:FSCan:SLISt:MARK:CLEar:SIGNal 20
:CALCulate:FSCan:SLISt:MARK:SIGNal
Syntax
:CALCulate:FSCan:SLISt:MARK:SIGNal <integer>
Description
Marks the specified signal.
Parameter
Name
Type
Range
Default
<integer>
Integer
1 to 2,000
--
Example
The following command marks Signal 20 in the signal table.
:CALCulate:FSCan:SLISt:MARK:SIGNal 20
:CALCulate:FSCan:SLISt:REPLace:METer
Syntax
:CALCulate:FSCan:SLISt:REPLace:METer <integer>
Description
Replaces the frequency of the currently selected signal with the currently set frequency of the meter.
Parameter
Name
Type
Range
Default
<integer>
Integer
1 to 2,000
--
Example
The following command replaces the frequency of Signal 20 with the frequency of the meter.
:CALCulate:FSCan:SLISt:REPLace:METer 20
:CALCulate:FSCan:SLISt:SET:METer
Syntax
:CALCulate:FSCan:SLISt:SET:METer <integer>
Description
Sets the frequency of the meter to the frequency of the specified signal.