Instructions
Chapter 3 Command System for EMI Mode  RIGOL 
RSA3000E Programming Guide  3-9 
Parameter 
Name 
Type 
Range 
Default 
<n>  
Discrete 
1|2|3|4|5|6|7|8 
-- 
Remarks 
When no peak is found, a prompt message "No peak found" is displayed on the screen. 
Example 
The following command performs one left peak search, and marks with Marker 2.  
:CALCulate:FSCan:MARKer2:MAXimum:LEFT 
:CALCulate:FSCan:MARKer<n>:MAXimum:NEXT 
Syntax 
:CALCulate:FSCan:MARKer<n>:MAXimum:NEXT
Description
Searches for and marks the peak whose amplitude on the trace is next to that of the current peak and 
which meets the peak search condition. 
Parameter 
Name 
Type 
Range 
Default 
<n>  
Discrete 
1|2|3|4|5|6|7|8 
-- 
Remarks
When no peak is found, a prompt message "No peak found" is displayed on the screen. 
Example 
The following command performs one next peak search, and marks with Marker 2.  
:CALCulate:FSCan:MARKer2:MAXimum:NEXT 
:CALCulate:FSCan:MARKer<n>:MAXimum:RIGHt 
Syntax 
:CALCulate:FSCan:MARKer<n>:MAXimum:RIGHt
Description
Searches for and marks the nearest peak which is located at the right side of the current peak and meets 
the peak search condition.
Parameter
Name 
Type 
Range 
Default 
<n>  
Discrete 
1|2|3|4|5|6|7|8 
-- 
Remarks
When no peak is found, a prompt message "No peak found" is displayed on the screen. 
Example 
The following command performs one right peak search, and marks with Marker 2.  
:CALCulate:FSCan:MARKer2:MAXimum:RIGHt 