Instructions
RIGOL Chapter 3 Command System for EMI Mode
3-8 RSA3000E Programming Guide
:CALCulate:FSCan:MARker<n>:FUNCtion:MAMarker?
Syntax
:CALCulate:FSCan:MARker<n>:FUNCtion:MAMarker?
Description
Queries the result measured at the marker.
Parameter
Name
Type
Range
Default
<n>
Discrete
1
|
2|3
|
4|5|6|7|8
--
Return Format
The query returns the result measured at the marker in scientific notation, including detector 1 amplitude
value, detector 2 amplitude value, detector 3 amplitude value and frequency value at the marker. The
return values are separated by commas. If the detector is disabled or the measurement is not completed,
the query returns -1.700000000e+02.
Example
The following command returns the result measured at Marker 1
-1.700000000e+02,-1.700000000e+02,-1.700000000e+02,5.150000000e+08.
:CALCulate:FSCan:MARker1:FUNCtion:MAMarker?
:CALCulate:FSCan:MARKer<n>:MAXimum
Syntax
:CALCulate:FSCan:MARKer<n>:MAXimum
Description
Performs one peak search based on the search mode set by
the
:CALCulate:FSCan:MARKer:PEAK:SEARch:MODE command and marks it with the specified marker.
Parameter
Name
Type
Range
Default
<n>
Discrete
1|2|3|4|5|6|7|8
--
Remarks
When no peak is found, a prompt message "No peak found" is displayed on the screen.
Example
The following command performs one peak search, and marks with Marker 2.
:CALCulate:FSCan:MARKer2:MAXimum:MAX
:CALCulate:FSCan:MARKer<n>:MAXimum:LEFT
Syntax
:CALCulate:FSCan:MARKer<n>:MAXimum:LEFT
Description
Searches for and marks the nearest peak which is located at the left side of the current peak and meets the
peak search condition.