Instructions

Chapter 3 Command System for EMI Mode RIGOL
RSA3000E Programming Guide 3-5
Parameter
Name
Type
Range
Default
--
Bool
OFF|ON|0|1
OFF|0
Remarks
When you enable the couple marker function, moving any marker will enable other markers (except the Off
marker) to move with it.
Return Format
The query returns 0 or 1.
Example
The following command disables the couple marker function.
:CALCulate:FSCan:MARKer:COUPle:STATe OFF or :CALCulate:FSCan:MARKer:COUPle:STATe 0
The following query returns 0.
:CALCulate:FSCan:MARKer:COUPle:STATe?
:CALCulate:FSCan:MARKer:PEAK:EXCursion
Syntax
:CALCulate:FSCan:MARKer:PEAK:EXCursion <rel_ampl>
:CALCulate:FSCan:MARKer:PEAK:EXCursion?
Description
Sets the peak excursion. The unit is dB.
Queries the peak excursion.
Parameter
Name
Type
Range
Default
<rel_ampl>
Consecutive Real Number
0 dB to 100 dB
6 dB
Return Format
The query returns the peak excursion in scientific notation.
Example
The following command sets the peak excursion to 12 dB.
:CALCulate:FSCan:MARKer:PEAK:EXCursion 12
The following query returns 1.200000000e+01.
:CALCulate:FSCan:MARKer:PEAK:EXCursion?
:CALCulate:FSCan:MARKer:PEAK:EXCursion:STATe
Syntax
:CALCulate:FSCan:MARKer:PEAK:EXCursion:STATe OFF|ON|0|1
:CALCulate:FSCan:MARKer:PEAK:EXCursion:STATe?
Description
Enables or disables the peak excursion function.
Queries the status of the peak excursion function.