Instructions
Chapter 2 Command Systems for GPSA and RTSA Modes RIGOL
RSA3000E Programming Guide 2-209
:TRIGger:SEQuence:EXTernal1:DELay:STATe ON or :TRIGger:SEQuence:EXTernal1:DELay:STATe 1
The following query returns 1.
:TRIGger:SEQuence:EXTernal1:DELay:STATe?
:TRIGger[:SEQuence]:EXTernal<n>:SLOPe
Syntax
:TRIGger:SEQuence:EXTernal<n>:SLOPe POSitive|NEGative
:TRIGger:SEQuence:EXTernal<n>:SLOPe?
Description
Sets the trigger edge for the external trigger.
Queries the trigger edge for the external trigger.
Parameter
Name
Type
Range
Default
<n>
Discrete
1|2
——
——
Keyword
POSitive|NEGative
POSitive
Remarks
When the parameter n is set to 1, it indicates External Trigger 1; when set to 2, it indicates External Trigger
2.
POSitive: indicates the rising edge.
NEGative: indicates the falling edge.
Return Format
The query returns POS or NEG.
Example
The following command sets the trigger edge of External Trigger 1 to Positive.
:TRIGger:SEQuence:EXTernal1:SLOPe POSitive
The following query returns POS.
:TRIGger:SEQuence:EXTernal1:SLOPe?
:TRIGger[:SEQuence]:FMT:APTRigger
Syntax
:TRIGger[:SEQuence]:FMT:APTRigger <number>
:TRIGger[:SEQuence]:FMT:APTRigger?
Description
Sets the number of times for acquisition after each effective trigger is completed.
Queries the number of times for acquisition after each effective trigger is completed.
Parameter
Name
Type
Range
Default
<number>
Integer
Refer to "Remarks"
1
Remarks
The minimum value of the parameter <number> is 1. When the trigger criteria of FMT is "Inside" or