Instructions

RIGOL Chapter 2 Command Systems for GPSA and RTSA Modes
2-176 RSA3000E Programming Guide
Remarks
This command is only valid when the T-power measurement in GPSA mode is enabled.
Return Format
The query returns the stop line in scientific notation. The unit is s.
Example
The following command sets the stop line to 10 ms.
:SENSe:TPOWer:RLIMit 0.01
The following query returns 1.000000000e-02.
:SENSe:TPOWer:RLIMit?
[:SENSe]:VSWR:FREFlect
Syntax
[:SENSe]:VSWR:FREFlect
Description
When you send this command, the analyzer performs the first measurement for VSWR mesurement
(disconnect the DUT).
Remarks
This command is only valid when the VSWR mesurement is enabled in GPSA mode.
Before sending this command, disconnect the DUT.
[:SENSe]:VSWR:NREFlect
Syntax
[:SENSe]:VSWR:NREFlect
Description
When you send this command, the analyzer performs the second measurement for VSWR mesurement.
Remarks
This command is only valid when the VSWR mesurement is enabled in GPSA mode.
Before sending this command, disconnect the DUT and send the [:SENSe]:VSWR:FREFlect command, and
then connect the DUT.
[:SENSe]:VSWR:RESet
Syntax
[:SENSe]:VSWR:RESet
Description
Resets the VSWR measurement parameters to their default values.
Remarks
This command is only valid when the VSWR mesurement is enabled in GPSA mode.