Instructions

RIGOL Chapter 2 Command Systems for GPSA and RTSA Modes
2-174 RSA3000E Programming Guide
[:SENSe]:TPOWer:AVERage:TCONtrol
Syntax
[:SENSe]:TPOWer:AVERage:TCONtrol EXPonential|REPeat
[:SENSe]:TPOWer:AVERage:TCONtrol?
Description
Sets the average mode of the T-power measurement.
Queries the average mode of the T-power measurement.
Parameter
Name
Type
Range
Default
——
Keyword
EXPonential|REPeat
EXPonential
Remarks
EXPonential: indicates the exponential average.
REPeat: indicates the repeat average.
When "EXPonential" is selected, the result is the exponential average of the measurement results obtained
in the past N times (N is specified in "
[:SENSe]:TPOWer:AVERage:COUNt").
When "REPeat" is selected, the result is the arithmetic average of the measurement results obtained in the
past N times (N is specified in "
[:SENSe]:TPOWer:AVERage:COUNt").
This command is only valid when the T-power measurement in GPSA mode is enabled.
Return Format
The query returns EXP or REP.
Example
The following command sets the average mode to Repeat.
:SENSe:TPOWer:AVERage:TCONtrol REPeat
The following query returns REP.
:SENSe:TPOWer:AVERage:TCONtrol?
[:SENSe]:TPOWer:LLIMit
Syntax
[:SENSe]:TPOWer:LLIMit <time>
[:SENSe]:TPOWer:LLIMit?
Description
Sets the start line for the T-power measurement.
Queries the start line for the T-power measurement.
Parameter
Name
Type
Range
Default
<time>
Consecutive Real Number
0 μs to (current value of stop line)
0 μs
Remarks
This command is only valid when the T-power measurement in GPSA mode is enabled.
Return Format
The query returns the start line in scientific notation. The unit is s.