Instructions
Chapter 2 Command Systems for GPSA and RTSA Modes  RIGOL 
RSA3000E Programming Guide  2-171 
Return Format 
The query returns the average count in integer.
Example
The following command sets the average count to 100.  
:SENSe:TOI:AVERage:COUNt 100 
The following query returns 100.   
:SENSe:TOI:AVERage:COUNt? 
[:SENSe]:TOI:AVERage[:STATe] 
Syntax 
[:SENSe]:TOI:AVERage[:STATe] OFF|ON|0|1  
[:SENSe]:TOI:AVERage[:STATe]? 
Description
Enables or disables the average measurement function of the TOI measurement. 
Queries the status of the average measurement function of the TOI measurement. 
Parameter
Name 
Type 
Range 
Default 
—— 
Bool 
OFF|ON|0|1 
ON|1 
Remarks 
This command is only valid when the TOI measurement in GPSA mode is enabled. 
Return Format
The query returns 0 or 1. 
Example
The following command enables the average measurement function.   
:SENSe:TOI:AVERage:STATe ON or :SENSe:TOI:AVERage:STATe 1 
The following query returns 1.   
:SENSe:TOI:AVERage:STATe? 
[:SENSe]:TOI:AVERage:TCONtrol 
Syntax 
[:SENSe]:TOI:AVERage:TCONtrol EXPonential|REPeat  
[:SENSe]:TOI:AVERage:TCONtrol? 
Description
Selects the average mode of the TOI measurement.  
Queries the average mode of the TOI measurement. 
Parameter
Name 
Type 
Range 
Default 
—— 
Keyword 
EXPonential|REPeat 
EXPonential 