Instructions
Chapter 2 Command Systems for GPSA and RTSA Modes RIGOL
RSA3000E Programming Guide 2-171
Return Format
The query returns the average count in integer.
Example
The following command sets the average count to 100.
:SENSe:TOI:AVERage:COUNt 100
The following query returns 100.
:SENSe:TOI:AVERage:COUNt?
[:SENSe]:TOI:AVERage[:STATe]
Syntax
[:SENSe]:TOI:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:TOI:AVERage[:STATe]?
Description
Enables or disables the average measurement function of the TOI measurement.
Queries the status of the average measurement function of the TOI measurement.
Parameter
Name
Type
Range
Default
——
Bool
OFF|ON|0|1
ON|1
Remarks
This command is only valid when the TOI measurement in GPSA mode is enabled.
Return Format
The query returns 0 or 1.
Example
The following command enables the average measurement function.
:SENSe:TOI:AVERage:STATe ON or :SENSe:TOI:AVERage:STATe 1
The following query returns 1.
:SENSe:TOI:AVERage:STATe?
[:SENSe]:TOI:AVERage:TCONtrol
Syntax
[:SENSe]:TOI:AVERage:TCONtrol EXPonential|REPeat
[:SENSe]:TOI:AVERage:TCONtrol?
Description
Selects the average mode of the TOI measurement.
Queries the average mode of the TOI measurement.
Parameter
Name
Type
Range
Default
——
Keyword
EXPonential|REPeat
EXPonential