Instructions
RIGOL Chapter 2 Command Systems for GPSA and RTSA Modes
2-170 RSA3000E Programming Guide
Example
The following command enables the auto sweep time.
:SENSe:SWEep:TIME:AUTO ON or :SENSe:SWEep:TIME:AUTO 1
The following query returns 1.
:SENSe:SWEep:TIME:AUTO?
[:SENSe]:SWEep:TIME:AUTO:RULes
Syntax
[:SENSe]:SWEep:TIME:AUTO:RULes NORMal|ACCuracy
[:SENSe]:SWEep:TIME:AUTO:RULes?
Description
Selects the sweep type.
Queries the sweep type.
Parameter
Name
Type
Range
Default
——
Keyword
NORMal|ACCuracy
NORMal
Remarks
NORMal: indicates normal.
ACCuracy: indicates accuracy.
This command is only available for GPSA mode
Return Format
The query returns NORM or ACC.
Example
The following command selects the sweep type to Accuracy.
:SENSe:SWEep:TIME:AUTO:RULes ACCuracy
The following query returns ACC.
:SENSe:SWEep:TIME:AUTO:RULes?
[:SENSe]:TOI:AVERage:COUNt
Syntax
[:SENSe]:TOI:AVERage:COUNt <integer>
[:SENSe]:TOI:AVERage:COUNt?
Description
Sets the average count of the TOI measurement.
Queries the average count of the TOI measurement.
Parameter
Name
Type
Range
Default
<integer>
Integer
1 to 1,000
10