Instructions
Chapter 2 Command Systems for GPSA and RTSA Modes  RIGOL 
RSA3000E Programming Guide  2-165 
Parameter 
Name 
Type 
Range 
Default 
-- 
Bool 
OFF|ON|0|1 
OFF|0 
Remarks 
This command is only available for RTSA mode and valid when the SSC function is enabled. 
Return Format
The query returns 0 or 1. 
Example
The following command enables Marker 2. 
:SENSe:SIGCapture:2FSK:MARK2:SWitch:STATe ON or :SENSe:SIGCapture:2FSK:MARK2:SWitch:STATe 1 
The following query returns 1. 
:SENSe:SIGCapture:2FSK:MARK2:SWitch:STATe? 
[:SENSe]:SIGCapture:2FSK:MAXHold[:STATe] 
Syntax 
[:SENSe]:SIGCapture:2FSK:MAXHold[:STATe] OFF|ON|0|1 
[:SENSe]:SIGCapture:2FSK:MAXHold[:STATe]? 
Description
Enables or disables the Max Hold. 
Queries the on/off status of the Max Hold.
Parameter
Name 
Type 
Range 
Default 
-- 
Bool 
OFF|ON|0|1 
OFF|0 
Remarks 
When the Max Hold is enabled (i.g. the SSC function is enabled), the signal captured each time will be 
compared with the last captured signal, whichever the larger value will be displayed as the max hold trace. 
Return Format
The query returns 0 or 1. 
Example
The following command enables the Max Hold. 
:SENSe:SIGCapture:2FSK:MAXHold:STATe ON or :SENSe:SIGCapture:2FSK:MAXHold:STATe 1 
The following query returns 1. 
:SENSe:SIGCapture:2FSK:MAXHold:STATe? 
[:SENSe]:SIGCapture:2FSK:PEAKAmp? 
Syntax 
[:SENSe]:SIGCapture:2FSK:PEAKAmp? <peaknum> 
Description 
Queries the amplitude of the nth peak in the SSC measurement results. 










