Instructions
RIGOL  Chapter 2 Command Systems for GPSA and RTSA Modes 
2-154  RSA3000E Programming Guide 
[:SENSe]:MCHPower:AVERage:TCONtrol 
Syntax 
[:SENSe]:MCHPower:AVERage:TCONtrol EXPonential|REPeat  
[:SENSe]:MCHPower:AVERage:TCONtrol? 
Description
Selects the average mode of the multi-channel power measurement.  
Queries the average mode of the multi-channel power measurement.
Parameter
Name 
Type 
Range 
Default 
—— 
Keyword 
EXPonential|REPeat 
EXPonential 
Remarks
EXPonential: indicates the exponential average. 
REPeat: indicates the repeat average. 
When "EXPonential" is selected, the result is the exponential average of the measurement results obtained 
in the past N times (N is specified in 
[:SENSe]:MCHPower:AVERage:COUNt).  
When "REPeat" is selected, the result is the arithmetic average of the measurement results obtained in the 
past N times (N is specified in 
[:SENSe]:MCHPower:AVERage:COUNt). 
This command is only valid when the multi-channel power measurement in GPSA mode is enabled.   
Return Format 
The query returns EXP or REP
. 
Example 
The following command sets the average mode to Repeat.  
:SENSe:MCHPower:AVERage:TCONtrol REPeat 
The following query returns REP.   
:SENSe:MCHPower:AVERage:TCONtrol? 
[:SENSe]:MCHPower:FREQuency:CENTer 
Syntax 
[:SENSe]:MCHPower:FREQuency:CENTer <freq>  
[:SENSe]:MCHPower:FREQuency:CENTer?
Description 
Sets the center frequency of the selected channel in the channel editing table.  
Queries the center frequency of the selected channel in the channel editing table.
Parameter
Name 
Type 
Range 
Default 
<freq>  
Consecutive Real Number 
5 Hz to 3 GHz 
1.5 GHz 
Remarks 
This command is only valid when the multi-channel power measurement is enabled in GPSA mode. 
Return Format
The query returns the center frequency of the selected channel in scientific notation. The unit is Hz. 