Instructions

RIGOL Chapter 2 Command Systems for GPSA and RTSA Modes
2-150 RSA3000E Programming Guide
Remarks
This command is invalid when the advance measurement function (AMK) is enabled in GPSA mode or when
in RTSA mode.
[:SENSe]:HDISt:AVERage:COUNt
Syntax
[:SENSe]:HDISt:AVERage:COUNt <integer>
[:SENSe]:HDISt:AVERage:COUNt?
Description
Sets the average count of the harmonic distortion measurement.
Queries the average count of the harmonic distortion measurement.
Parameter
Name
Type
Range
Default
<integer>
Integer
1 to 1,000
10
Remarks
This command is only valid when the harmonic distortion measurement in GPSA mode is enabled.
Return Format
The query returns the average count in integer.
Example
The following command sets the average count to 100.
:SENSe:HDISt:AVERage:COUNt 100
The following query returns 100.
:SENSe:HDISt:AVERage:COUNt?
[:SENSe]:HDISt:AVERage[:STATe]
Syntax
[:SENSe]:HDISt:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:HDISt:AVERage[:STATe]?
Description
Enables or disables the average measurement function of the harmonic distortion measurement.
Queries the status of the average measurement function of the harmonic distortion measurement.
Parameter
Name
Type
Range
Default
——
Bool
OFF|ON|0|1
OFF|0
Remarks
This command is only valid when the harmonic distortion measurement in GPSA mode is enabled.
Return Format
The query returns 0 or 1.