Instructions
Chapter 2 Command Systems for GPSA and RTSA Modes  RIGOL 
RSA3000E Programming Guide  2-149 
Parameter 
Name 
Type 
Range 
Default 
<freq> 
Consecutive Real 
Number 
GPSA mode 
0 Hz to Fmax (zero span) 
0 Hz to (Fmax - 10 Hz) 
(non-zero span) 
center 
frequency-span/2 
RTSA mode 
0 Hz to (Fmax - 5 kHz) 
Return Format
The query returns the start frequency in scientific notation. The unit is Hz. 
Example
The following command sets the start frequency to 100 MHz. 
:SENSe:FREQuency:STARt 100000000 
The following query returns 1.000000000e+08. 
:SENSe:FREQuency:STARt? 
[:SENSe]:FREQuency:STOP 
Syntax   
[:SENSe]:FREQuency:STOP <freq>  
[:SENSe]:FREQuency:STOP? 
Description
Sets the stop frequency.  
Queries the stop frequency.
Parameter
Name 
Type 
Range 
Default 
<freq> 
Consecutive Real 
Number 
GPSA mode 
0 Hz to Fmax (zero span) 
10 Hz to Fmax (non-zero 
span) 
center 
frequency+span/2 
RTSA mode 
5 kHz to Fmax 
Return Format
The query returns the stop frequency in scientific notation. The unit is Hz. 
Example
The following command sets the stop frequency to 10 MHz. 
:SENSe:FREQuency:STOP 10000000 
The following query returns 1.000000000e+07. 
:SENSe:FREQuency:STOP? 
[:SENSe]:FREQuency:TUNE:IMMediate 
Syntax   
[:SENSe]:FREQuency:TUNE:IMMediate 
Description
Automatically searches for the signal within the full frequency range, and adjusts the frequency and 
amplitude for optimal display effect of the signal. 