Instructions
RIGOL  Chapter 2 Command Systems for GPSA and RTSA Modes 
2-140  RSA3000E Programming Guide 
Example 
The following command sets the average mode to Repeat.  
:SENSe:EBWidth:AVERage:TCONtrol REPeat 
The following query returns REP.   
:SENSe:EBWidth:AVERage:TCONtrol? 
[:SENSe]:EBWidth:FREQuency:SPAN 
Syntax   
[:SENSe]:EBWidth:FREQuency:SPAN <freq>  
[:SENSe]:EBWidth:FREQuency:SPAN? 
Description
Sets the span of the emission bandwidth measurement. 
Queries the span of the emission bandwidth measurement. 
Parameter
Name 
Type 
Range 
Default 
<freq> 
Consecutive Real Number 
10 Hz to 3 GHz 
2 MHz 
Remarks
This command is only valid when the emission bandwidth measurement in GPSA mode is enabled.  
This setting changes the span of the analyzer. 
Return Format
The query returns the span in scientific notation. Its unit is Hz. 
Example
The following command sets the span to 10 MHz. 
:SENSe:EBWidth:FREQuency:SPAN 10000000 
The following query returns 1.000000000e+07.  
:SENSe:EBWidth:FREQuency:SPAN? 
[:SENSe]:EBWidth:MAXHold:STATe 
Syntax   
[:SENSe]:EBWidth:MAXHold:STATe OFF|ON|0|1  
[:SENSe]:EBWidth:MAXHold:STATe? 
Description
Enables or disables the EBW Max Hold.   
Queries the status of the EBW Max Hold. 
Parameter
Name 
Type 
Range 
Default 
—— 
Bool 
OFF|ON|0|1 
OFF|0 
Remarks 
This command is only valid when the emission bandwidth measurement in GPSA mode is enabled.  
When Max Hold is enabled, each measurement result is compared with the previous result, and then display 