Instructions
RIGOL  Chapter 2 Command Systems for GPSA and RTSA Modes 
2-138  RSA3000E Programming Guide 
[:SENSe]:DETector:TRACe<n>:AUTO 
Syntax 
[:SENSe]:DETector:TRACe<n>:AUTO OFF|ON|0|1 
[:SENSe]:DETector:TRACe<n>:AUTO? 
Description
Enables or disables the Detector Auto function for the specified trace.  
Queries the status of the Detector Auto function for the specified trace.
Parameter
Name 
Type 
Range 
Default 
<n> 
Discrete 
1|2|3|4|5|6 
1 
—— 
Bool 
OFF|ON|0|1 
ON|1 
Return Format
The query returns 0 or 1. 
Example
The following command enables the Detector Auto function for Trace 1. 
:SENSe:DETector:TRACe1:AUTO ON or :SENSe:DETector:TRACe1:AUTO 1 
The following query returns 1.  
:SENSe:DETector:TRACe1:AUTO? 
[:SENSe]:EBWidth:AVERage:COUNt 
Syntax 
[:SENSe]:EBWidth:AVERage:COUNt <integer>   
[:SENSe]:EBWidth:AVERage:COUNt?
Description 
Sets the average count of the emission bandwidth measurement.  
Queries the average count of the emission bandwidth measurement.
Parameter
Name 
Type 
Range 
Default 
<integer> 
Integer 
1 to 1,000 
10 
Remarks
This command is only valid when the emission bandwidth measurement in GPSA mode is enabled. 
Return Format
The query returns the average count in integer. 
Example
The following command sets the average count to 100.  
:SENSe:EBWidth:AVERage:COUNt 100 
The following query returns 100.  
:SENSe:EBWidth:AVERage:COUNt? 