Instructions
Chapter 2 Command Systems for GPSA and RTSA Modes  RIGOL 
RSA3000E Programming Guide  2-65 
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel 
Syntax 
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel <ampl> 
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel?
Description
Sets the reference level. 
Queries the reference level. 
Parameter
Name 
Type 
Range 
Default 
<ampl> 
Consecutive Real Number 
-170 dBm to 30 dBm 
0 dBm 
Return Format
The query returns the reference level in scientific notation. 
Example
The following command sets the reference level to -10 dB.  
:DISPlay:WINDow:TRACe:Y:SCALe:RLEVel -10 
The following query returns -1.000000e+01.   
:DISPlay:WINDow:TRACe:Y:SCALe:RLEVel? 
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel:OFFSet 
Syntax 
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel:OFFSet <rel_ampl> 
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel:OFFSet?
Description
Sets the reference level offset. 
Queries the reference level offset. 
Parameter
Name 
Type 
Range 
Default 
<rel_ampl> 
Consecutive Real Number 
-300 dB to 300 dB 
0 dB 
Remarks
The offset of the reference level only modifies the reference level and amplitude readout of the marker, but 
does not change the position of the curve.
Return Format
The query returns the offset of the reference level in scientific notation. 
Example
The following command sets the reference level offset to 10 dB.  
:DISPlay:WINDow:TRACe:Y:SCALe:RLEVel:OFFSet 10 
The following query returns 1.000000e+01.   
:DISPlay:WINDow:TRACe:Y:SCALe:RLEVel:OFFSet? 