Instructions
Chapter 2 Command Systems for GPSA and RTSA Modes  RIGOL 
RSA3000E Programming Guide  2-21 
:CALCulate:MARKer<n>:FCOunt[:STATe] 
Syntax
:CALCulate:MARKer<n>:FCOunt[:STATe] OFF|ON|0|1 
:CALCulate:MARKer<n>:FCOunt[:STATe]? 
Description
Enables or disables the frequency counter function of the specified marker.   
Queries the status of the frequency counter function of the specified marker. 
Parameter
Name 
Type 
Range 
Default 
<n> 
Discrete 
1|2|3|4|5|6|7|8 
—— 
—— 
Bool 
OFF|ON|0|1 
OFF|0 
Remarks 
This command is only available for GPSA mode. 
Return Format
The query returns 0 or 1. 
Example
The following command enables the frequency counter function of Marker 2.   
:CALCulate:MARKer2:FCOunt:STATe ON or :CALCulate:MARKer2:FCOunt:STATe 1 
The following query returns 1.  
:CALCulate:MARKer2:FCOunt:STATe?  
:CALCulate:MARKer<n>:FCOunt:X? 
Syntax
:CALCulate:MARKer<n>:FCOunt:X? 
Description
Queries the readout of the frequency counter of the specified marker. 
Parameter
Name 
Type 
Range 
Default 
<n> 
Discrete 
1|2|3|4|5|6|7|8 
—— 
Remarks 
This command is only available for GPSA mode. 
Return Format
The query returns the readout in integer. Its unit is Hz. 
When the frequency counter function is disabled, the command returns NAN. 