Instructions
RIGOL  Chapter 4 Command System for VSA Mode 
4-70  RSA3000E Programming Guide 
The following query returns 1.000000000e-01.  
:TRIGger:SEQuence:HOLDoff? 
:TRIGger[:SEQuence]:HOLDoff:STATe 
Syntax 
:TRIGger[:SEQuence]:HOLDoff:STATe OFF|ON|0|1 
:TRIGger[:SEQuence]:HOLDoff:STATe? 
Description 
Turns on or off the trigger holdoff function.   
Queries the status of the trigger holdoff function. 
Parameter 
Name 
Type 
Range 
Default 
-- 
Bool 
OFF|ON|0|1 
OFF|0 
Return Format 
The query returns 1 or 0. 
Example 
The following command enables the trigger holdoff function.  
:TRIGger:SEQuence:HOLDoff:STATe ON or :TRIGger:SEQuence:HOLDoff:STATe 1 
The following query returns 1. 
:TRIGger:SEQuence:HOLDoff:STATe? 
:TRIGger[:SEQuence]:LEVel 
Syntax 
:TRIGger[:SEQuence]:LEVel <ampl> 
:TRIGger[:SEQuence]:LEVel? 
Description 
Sets the trigger level of the IF power trigger. 
Queries the trigger level of the IF power trigger. 
Parameter 
Name 
Type 
Range 
Default 
<ampl> 
Consecutive Real Number 
-140 dBm to 30 dBm 
0 dBm 
Remarks 
This command is only valid when IF power trigger is selected. 
Return Format 
The query returns the trigger level in scientific notation. 
Example 
The following command sets the trigger level of IF power to 10 dBm. 
:TRIGger:SEQuence:LEVel 10 
The following query returns 1.000000e+01. 
:TRIGger:SEQuence:LEVel? 