Instructions
Chapter 4 Command System for VSA Mode  RIGOL 
RSA3000E Programming Guide  4-23 
:DISPlay:DDEMod:TRACe<n>:X[:SCALe]:RPOSition 
Syntax 
:DISPlay:DDEMod:TRACe<n>:X[:SCALe]:RPOSition LEFT|CENTer|RIGHt 
:DISPlay:DDEMod:TRACe<n>:X[:SCALe]:RPOSition? 
Description 
Sets the position from which the X-axis scale reference value is calculated for the specified trace. 
Queries the position from which the X-axis scale reference value is calculated for the specified trace. 
Parameter 
Name 
Type 
Range 
Default 
<n>  
Discrete 
1|2|3|4 
-- 
-- 
Keyword 
LEFT|CENTer|RIGHt 
LEFT 
Return Format
The query returns LEFT, CENT, or RIGH. 
Example
The following command sets the position from which the X-axis scale reference value is calculated for Trace 
1 to LEFT.  
:DISPlay:DDEMod:TRACe1:X:SCALe:RPOSition LEFT 
The following query returns LEFT. 
:DISPlay:DDEMod:TRACe1:X:SCALe:RPOSition? 
:DISPlay:DDEMod:TRACe<n>:X[:SCALe]:SPAN 
Syntax 
:DISPlay:DDEMod:TRACe<n>:X[:SCALe]:SPAN <real> 
:DISPlay:DDEMod:TRACe<n>:X[:SCALe]:SPAN? 
Description 
Sets the X-axis width of the selected trace. 
Queries the X-axis width of the selected trace. 
Parameter 
Name 
Type 
Range 
Default 
<n>  
Discrete 
1|2|3|4 
-- 
<real>  Consecutive Real Number 
Time-Domain Data: 0 to 
(1e+12) 
Frequency-Domain Data: 
0Hz to 1THz 
50 (38.2812 us) 
/3.125 MHz 
Remarks 
This command is valide only when the X-axis scale is set to Manual. 
When the time-domain data source is selected, the reference unit is symbol or s; when the 
frequency-domain data source is selected, the reference unit is Hz. 
Return Format
The query returns the X-axis width of the selected trace in scientific notation. 
Example
The following command sets the X-axis width of Trace 1 to 50.  