Instructions
RIGOL  Chapter 3 Command System for EMI Mode 
3-86  RSA3000E Programming Guide 
[:SENSe]:FSCan:TRACe:TYPE 
Syntax 
[:SENSe]:FSCan:TRACe:TYPE WRITe|AVERage|MAXHold|MINHold 
[:SENSe]:FSCan:TRACe:TYPE? 
Description 
Sets the type of the trace. 
Queries the type of the trace.
Parameter 
Name 
Type 
Range 
Default 
-- 
Keyword 
WRITe|AVERage|MAXHold|MINHold 
WRITe 
Remarks 
WRITe: indicates clear/write. 
AVERage: indicates average.
MAXHold: indicates the maximum hold. 
MINHold: indicates the minimum hold. 
Return Format
The query returns WRIT, AVER, MAXH, or MINH. 
Example
The following command sets the type of the current trace to MAXHold. 
:SENSe:FSCan:TRACe:TYPE MAXHold 
The following query returns MAXH. 
:SENSe:FSCan:TRACe:TYPE? 
[:SENSe]:METer:DETector:DWELl 
Syntax 
[:SENSe]:METer:DETector:DWELl <time> 
[:SENSe]:METer:DETector:DWELl? 
Description 
Sets the dwell time of the meter detector. 
Queries the dwell time of meter detector. 
Parameter 
Name 
Type 
Range 
Default 
<time> 
Consectutive Real Number 
10 ms to 100 s 
10 ms 
Return Format 
The query returns the dwell time in scientific notation. The unit is s. 
Example 
The following command sets the dwell time of the meter detector to 0.1 s. 
:SENSe:METer:DETector:DWELl 0.1 
The following query returns 1.000000000e-01. 
:SENSe:METer:DETector:DWELl? 