Instructions
RIGOL  Chapter 3 Command System for EMI Mode 
3-66  RSA3000E Programming Guide 
Example 
The following command adds two points to the current correction curve. 
:SENSe:CORRection:CSET1:DATA 900E6,0.3,1.0E9,0.35 
[:SENSe]:CORRection:CSET<n>:DELete 
Syntax 
[:SENSe]:CORRection:CSET<n>:DELete 
Description 
Deletes the specified amplitude correction setting. 
Parameter
Name 
Type 
Range 
Default 
<n> 
Discrete 
1|2|3|4 
-- 
Remarks 
<n>: 1 indicates antenna; 2 indicates cable; 3 indicates other, and 4 indicates user. 
Example 
The following command deletes the antenna correction setting. 
:SENSe:CORRection:CSET1:DELete 
[:SENSe]:CORRection:CSET<n>[:STATe] 
Syntax 
[:SENSe]:CORRection:CSET<n>[:STATe] OFF|ON|0|1 
[:SENSe]:CORRection:CSET<n>[:STATe]? 
Description 
Enables or disables the specified amplitude correction function. 
Queries the status of the specified amplitude correction function. 
Parameter 
Name 
Type 
Range 
Default 
<n> 
Discrete 
1|2|3|4 
-- 
-- 
Bool 
OFF|ON|0|1 
OFF|0 
Remarks 
<n>: 1 indicates antenna; 2 indicates cable; 3 indicates other, and 4 indicates user. 
Return Format 
The query returns 0 or 1. 
Example 
The following command disables the "Antenna" amplitude correction function. 
:SENSe:CORRection:CSET1:STATe OFF or :SENSe:CORRection:CSET1:STATe 0 
The following query returns 0. 
:SENSe:CORRection:CSET1:STATe? 