Instructions
Chapter 3 Command System for EMI Mode  RIGOL 
RSA3000E Programming Guide  3-63 
[:SENSe]:BANDwidth|BWIDth[:RESolution] 
Syntax 
[:SENSe]:BANDwidth|BWIDth[:RESolution] <freq> 
[:SENSe]:BANDwidth|BWIDth[:RESolution]?
Description 
Sets the resolution bandwidth (RBW) of the meter frequency.  
Queries the resolution bandwidth of the meter frequency. 
Parameter 
Name 
Type 
Range 
Default 
<freq>  
Discrete 
100 Hz to 10 MHz (at 1-3-10 step) 
120 kHz 
Remarks 
When the EMC measurement standard is "None" and "Gauss" filter is selected, the RBW range is as above.  
When "EMI" filter is selected, resolution bandwidth can only be 200 Hz, 9 kHz, 120 kHz, or 1 MHz. 
Return Format 
The query returns the resolution bandwidth in scientific notation. Its unit is Hz. 
Example 
The following command sets RBW to 1000 Hz. 
:SENSe:BANDwidth:RESolution 1000 
:SENSe:BWIDth:RESolution 1000 
The following query returns 1.000000000e+03. 
:SENSe:BANDwidth:RESolution? 
:SENSe:BWIDth:RESolution? 
[:SENSe]:BANDwidth|BWIDth[:RESolution]:AUTO 
Syntax
[:SENSe]:BANDwidth|BWIDth[:RESolution]:AUTO OFF|ON|0|1 
[:SENSe]:BANDwidth|BWIDth[:RESolution]:AUTO?
Description
Enables or disables the auto setting mode of the meter RBW.  
Queries the status of the auto setting mode of the meter RBW. 
Parameter
Name 
Type 
Range 
Default 
-- 
Bool 
OFF|ON|0|1 
ON|1 
Remarks
OFF: indicates "Manual" mode. You can use the [:SENSe]:BANDwidth|BWIDth[:RESolution] command to 
set the RBW value. 
ON: indicates "Auto" mode. The RBW automatically couple to the frequency of the meter. 
Frequency (Meter) Fmet 
RBW 
Fmet < 150 kHz  
200 Hz 
150 kHz ≤ Fmet < 30 MHz 
9 kHz 
30 MHz ≤ Fmet < 1 GHz 
120 kHz 
Fmet ≥ 1 GHz 
1 MHz 